Back to search

Article

Extreme Serum Titanium Concentration Induced by Catastrophic Acetabular Cup Failure: Unveiling a Unique Scenario of Titanium Alloy Debris Accumulation

2024-01-23

Abstract excerpt

The majority of contemporary total hip arthroplasty (THA) implants are constructed from Ti-alloys, which are generally believed to generate fewer adverse local tissue reactions (ALTR) compared to CoCr-alloys. This study presents a case of unusual primary THA failure where a substantial release of Ti-alloy debris was observed. A 52-year-old active male underwent THA after posttraumatic aseptic necrosis of the femor...

Topics

Open a Topic to create a Post that cites this publication.

Identifiers and source

Literature Corpus work
3c427f82-bfeb-57c8-8a81-25f529e205d0
DOI
10.20944/preprints202401.1635.v1
Open publication

Related research

Semantic proximity does not establish scientific evidence.

Click a neighbor to travelStep 1 · 12 closest
Interactive article relationship graphSelect a related publication card to move it into the centre and load its closest explainable connections. Solid lines are source-backed structured connections. Dashed lines are semantic discovery signals and are not scientific evidence.
Extreme Serum Titanium Concentration Induced by Catastrophic Acetabular Cup Failure: Unveiling a Unique Scenario of Titanium Alloy Debris AccumulationDOI 10.20944/preprints202401.1635.v1
Select a neighboring publication to make it the new centre.