Does the Capsule Interfere with Performance of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Identification of Cryptococcus neoformans and Cryptococcus gattii?
Journal of clinical microbiology - 1 Feb 2016
Thomaz Danilo Y, Grenfell Rafaella C, Vidal Monica S M, Giudice Mauro C, Del Negro Gilda M B, Juliano Luiz, Benard Gil, de Almeida Júnior João N
Abstract excerpt
We described the impact of the capsule size for Cryptococcus neoformans and Cryptococcus gattii identification at the species level by Bruker matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS). After experimental capsule size modulation, we observed that reducing the capsule size resulted in improved identification by Bruker MALDI-TOF MS across all of the reference strains...