Article
Evaluation of the Roche COBAS AmpliPrep/COBAS TaqMan HCV Test.
Diagnostic microbiology and infectious disease - 1 Sept 2013
Pyne Michael T, Hillyard David R
Abstract excerpt
The performance of the Roche COBAS AmpliPrep/COBAS TaqMan HCV Test (CAP/CTM) was evaluated. The limit of detection was 9.5 IU/mL using the 3rd International Standard. Serial dilutions of each genotype demonstrated good reproducibility and linearity. Correlation with samples previously tested with the Roche Analyte Specific Reagent (ASR) was very good, with the CAP/CTM assay measuring 0.24 log IU/mL higher on...