Performance characteristics of the COBAS AMPLICOR hepatitis C virus MONITOR Test, version 2.0.
American journal of clinical pathology - 1 Aug 2000
Erali M, Ashwood E R, Hillyard D R
Abstract excerpt
We evaluated the performance characteristics of the COBAS AMPLICOR Hepatitis C Virus (HCV) MONITOR Test, version 2.0. Dilution studies using patient specimens demonstrated a lower limit of detection of 1,000 copies per milliliter. The assay was linear from 1,000 to 1 million HCV RNA copies per milliliter. Within-run precision and between-run precision were acceptable (approximately 0.100 and 0.14 SD for log10...