Article
A case report-Two manufacturers SAB testing kits can reveal different HLA antibody profiles-Identifying prozone and denatured antigen.
HLA - 1 Jul 2020
Battle Richard K, Henderson Lorna, Phelan Paul J, Latham Katy, Turner David M
Abstract excerpt
Recent reports have identified that the presence of non-native conformation HLA to which antibody can bind upon Luminex HLA Class I single antigen beads, can vary in levels between different manufacturers kits and that the prozone effect may also be specific to particular products. We present a case in which both prozone and non-native HLA reactive antibodies were observed, which raises important questions on how...