Article
Detection of Carbapenemases by Real-Time PCR and Melt Curve Analysis on the BD Max System
13 Feb 2014
Abstract excerpt
We developed a multiplex SYBR green real-time PCR for the BD Max instrument (BD Diagnostics, Sparks, MD) to detect a panel of carbapenemases. The assay was evaluated with 152 consecutive isolates sent to the German National Reference Laboratory, and 65/65 of the carbapenemase-positive and 87/87 of the carbapenemase-negative strains were identified correctly.