Phantom-based quantification of the spectral accuracy in dual-layer spectral CT for pediatric imaging at 100 kVp
2022-02-28
Abstract excerpt
<h4>Objectives</h4> To determine the spectral accuracy in detector-based dual-energy CT (DECT) at 100 kVp and wide (8 cm) collimation width for dose levels and object sizes relevant to pediatric imaging. <h4>Methods</h4> A spectral CT phantom containing tissue-equivalent materials and iodine inserts of varying concentrations was scanned on the latest generation detector-based DECT system. Two 3D-printed extension...