Impact of patient habitus and acquisition protocol on iodine quantification in dual source photon-counting CT

2022-12-20

Abstract excerpt

<h4>Objective</h4> Evaluation of iodine quantification accuracy with varying iterative reconstruction level, patient habitus, and acquisition mode on a first-generation dual-source photon-counting computed tomography (PCCT) system. <h4>Methods</h4> A multi-energy CT phantom (20 cm diameter/small) was imaged with and without an extension ring (30 by 40 cm/large). It was equipped with various iodine inserts (0.2, 0....