Article
Inadequate design of mutation detection panels prevents interpretation of variants of concern: results of an external quality assessment for SARS-CoV-2 variant detection.
Clinical chemistry and laboratory medicine - 27 Jan 2022
Buchta Christoph, Camp Jeremy V, Jovanovic Jovana, Radler Ulla, Benka Bernhard, Puchhammer-Stöckl Elisabeth, Müller Mathias M, Griesmacher Andrea, Aberle Stephan W, Görzer Irene
Abstract excerpt
OBJECTIVES: Mutation-specific PCR assays have quickly found their way into laboratory diagnostics due to their capacity to be a fast, easy to implement and high-throughput method for the detection of known SARS-CoV-2 variants of concern (VoCs). However, little is known about the performance of such assays in routine laboratory analysis. METHODS: The results reported in a recent round of an external quality...