Article
Validation of an Automated Detection Platform for Use with the Roche Linear Array Human Papillomavirus Genotyping Test
11 Sept 2008
Abstract excerpt
An automated platform (BeeBlot) was evaluated in parallel with the recommended protocol for the hybridization and detection steps of the Roche Linear Array human papillomavirus (HPV) genotyping test using DNA from 143 cervical specimens. Genotyping profiles showed 100% concordance between the methods, suggesting that automation could complement the Roche Linear Array for enhanced speed and reproducibility.
Read the complete abstract on PubMedTopics