Back to search

Article

Comparison of HyperArc, IMRT, and VMAT plans for Graves’ ophthalmopathy radiotherapy

2023-12-06

Abstract excerpt

<h4>Objective: </h4> To compare the dosimetric parameters of three different radiotherapy techniques, HyperArc (HA), volumetric modulated arc therapy (VMAT), and intensity-modulated radiotherapy (IMRT), for patients with Graves’ ophthalmopathy (GO), and to provide guidance for the clinical choice of radiotherapy methods for GO. Materials and Methods Eight patients with GO were enrolled in this study. HA, IMRT, an...

Topics

Open a Topic to create a Post that cites this publication.

Identifiers and source

Literature Corpus work
c1a8c757-cb33-575f-985c-db9698f2b793
DOI
10.21203/rs.3.rs-3681637/v1
Open publication

Related research

Semantic proximity does not establish scientific evidence.

Click a neighbor to travelStep 1 · 8 closest
Interactive article relationship graphSelect a related publication card to move it into the centre and load its closest explainable connections. Solid lines are source-backed structured connections. Dashed lines are semantic discovery signals and are not scientific evidence.
Comparison of HyperArc, IMRT, and VMAT plans for Graves’ ophthalmopathy radiotherapyDOI 10.21203/rs.3.rs-3681637/v1
Select a neighboring publication to make it the new centre.