Article
Age and lead configuration matter: A comparative study of RF-induced heating of epicardial and endocardial electronic devices in adult and pediatric anthropomorphic phantoms in 1.5 T MR
2022-11-04
Abstract excerpt
<h4>Background</h4> Children with congenital heart defects often have life-sustaining indications for a cardiac implantable electronic device (CIED). In children, these devices are typically sewn to the heart epicardium, but the FDA has never licensed an epicardial system as MR-Conditional due to limited data. Children’s hospitals default to either refusing MRI service to a vast majority of pediatric CIED patient...
Topics
Open a Topic to create a Post that cites this publication.
Identifiers and source
- Literature Corpus work
- 76d07a90-3258-5ed2-8a84-73faa08691fc
- DOI
- 10.1101/2022.11.03.515092