Back to search

Article

Quantitative Evaluation of apparent diffusion coefficient in a large multi-unit institution using the QIBA diffusion phantom

2020-09-11

Abstract excerpt

<h4>Purpose</h4> The purpose of this study was to determine the quantitative variability of diffusion weighted imaging and apparent diffusion coefficient values across a large fleet of MR systems. Using a NIST traceable magnetic resonance imaging diffusion phantom, imaging was reproducible and the measurements were quantitatively compared to known values. <h4>Methods</h4> A fleet of 23 clinical MRI scanners was in...

Topics

Open a Topic to create a Post that cites this publication.

Identifiers and source

Literature Corpus work
03781004-1ab5-5a83-9a4a-b7b52cc52511
DOI
10.1101/2020.09.09.20191403
Open publication

Related research

Semantic proximity does not establish scientific evidence.

Click a neighbor to travelStep 1 · 8 closest
Interactive article relationship graphSelect a related publication card to move it into the centre and load its closest explainable connections. Solid lines are source-backed structured connections. Dashed lines are semantic discovery signals and are not scientific evidence.
Quantitative Evaluation of apparent diffusion coefficient in a large multi-unit institution using the QIBA diffusion phantomDOI 10.1101/2020.09.09.20191403
Select a neighboring publication to make it the new centre.