Article
Evaluation of the automated Vitek 2 system for detection of various mechanisms of macrolide and lincosamide resistance in Staphylococcus aureus.
Journal of clinical microbiology - 1 Nov 2014
Filippin Lorenzo, Roisin Sandrine, Nonhoff Claire, Vandendriessche Stien, Heinrichs Amélie, Denis Olivier
Abstract excerpt
We evaluated the performance of the automated Vitek 2 system against disk diffusion for susceptibility testing of Staphylococcus aureus strains showing various resistance mechanisms to macrolides and lincosamides (ML). The Vitek 2 system showed 100% concordance with the D-zone test in detection of the most common resistance mechanisms to ML, including methylase and efflux systems.
Read the complete abstract on PubMedTopics
Share this publication in a Topic to start or enrich a Post.