Article
Outbreak of multidrug-resistant Pseudomonas aeruginosa infection following urodynamic studies traced to contaminated transducer.
The Journal of hospital infection - 1 Apr 2013
Bilavsky E, Pfeffer I, Tarabeia J, Schechner V, Abu-Hanna J, Grisaru-Soen G, Schwartz D, Navon-Venezia S, Carmeli Y
Abstract excerpt
We report a nosocomial outbreak of urinary tract infection by extremely drug resistant Pseudomonas aeruginosa, susceptible only to colistin. Infection in three patients followed urodynamic studies. Two of the three patients were children, one of whom also developed urosepsis. The investigation led to detection of contaminated pressure transducers. Genotyping confirmed that patient and transducer isolates were...