Is the sample matrix suppressing target detection?

by Ruth W.

An ELISA reports little or no target in tissue lysate, while the standard curve and positive control are acceptable. Which control best separates matrix-dependent signal suppression from genuinely low target abundance: serial dilution of the lysate, purified-target spike recovery in the lysate, or both? What recovery or dilution pattern would identify matrix interference?

1
Safety · report, block, mute

Blocking hides the author in your feeds and prevents direct replies between you. Muting hides a Topic. Public posts remain public.

Crosspost to another Topic

Write your own title and commentary. The original is linked, not copied. To crosspost a crosspost, open its original first.

0
CipherRaven79

First confirm that the lysate dilutions and spikes fall within the calibrated range. Use both controls: serial dilution tests whether the measured concentration becomes proportional after reducing matrix load, while a purified-target spike tests recovery in the same lysate. Increasing apparent concentration after dilution, once corrected for the dilution factor, supports matrix-dependent suppression; poor spike recovery that improves with dilution strengthens that interpretation. Persistently acceptable spike recovery and dilutional parallelism would make low endogenous abundance more plausible.